Products > Respiratory Care > Respiratory Humidifier > CH580
CH580
Respiratory Humidifier
CH580
Respiratory Humidifier
The CH580® Respiratory Humidifier is designed to warm and add humidity to gases, which are delivered to the patient requiring artificial respiration using a respiratory care equipment, such as ventilators, BiPAPs, CPAPs, etc., where the gas flow remains at constant level.
The CH580® can be used on patients of all age groups, requiring any kind of invasive OR non-invasive artificial respiration.
The gas flows through heated water surface and then warmed and humidified, based on the mode setting selected (invasive mode / non-invasive mode) on the front control panel. The temperature of the gas delivered to the patient (patient end temperature) is monitored using the temperature probe and displayed on the unit. The patient end temperature and chamber temperature are controlled by humidifier based on the mode setting selected.
The CH580® Respiratory Humidifier is designed to warm and add humidity to gases, which are delivered to the patient requiring artificial respiration using a respiratory care equipment, such as ventilators, BiPAPs, CPAPs, etc., where the gas flow remains at constant level.
The CH580® can be used on patients of all age groups, requiring any kind of invasive OR non-invasive artificial respiration.
The gas flows through heated water surface and then warmed and humidified, based on the mode setting selected (invasive mode / non-invasive mode) on the front control panel. The temperature of the gas delivered to the patient (patient end temperature) is monitored using the temperature probe and displayed on the unit. The patient end temperature and chamber temperature are controlled by humidifier based on the mode setting selected.
Salient Features
- Compact, Light Weight & Easy to Operate
- Servo Controlled with Built-in Overheat Protection
- With Heater-Wire Functionality
- Easy to Read Chamber Temperature & Patient’s Airway Temperature Display and Alarms
- Adaptable with all Respirators
Specifications
Device Characteristics
| Thermal Cutout | 110 + 6 °C |
| Warm-up Period | < 30 minutes |
| Input Flow Range | Upto 40 lpm with humidity output > 33mg/L (Invasive Mode) Upto 80 lpm with humidity output > 10mg/L (Non-Invasive Mode) |
| Display | Three Digit, 14 mm, 7 Segment Green LED Display |
| Temperature Display Range | 10 ~ 70 °C |
| Temperature Display Resolution | + 0.1 °C |
| Temperature Display Accuracy | + 0.5 °C (in the temperature range of 25~45 °C) |
| Default Temperature Display | Lower of the Airway / Chamber Temperature |
| Alarms | High Temperature Alarm, Low Temperature Alarm, Probe Disconnect / Sensor Failure Alarm |
Electrical Characteristics
| Power Supply | 230 VAC ± 10%, 50/60 Hz, Sine Wave |
| Heater Plate Capacity | 150 W (at nominal mains voltage) |
| Heater Wire Rating | 22 + 5 VAC; 30 W; 1.36 Amp (Max) |
| Electrical Fuse | 2 x 2 Amp, 250V, Fast blow (for Main Unit); 1 x 4 Amp, 250V, Fast Blow (for Heater-Wire) |
Physical Characteristics
| Dimensions (L x W x H) | 165 x 135 x 155 mm³ |
| Weight | 3.0 kg |
Environmental Characteristics
| Temperature Range | Operation: 18 ~ 26 °C Storage: -10 ~ 50 °C |
| Relative Humidity Range | Operation: Upto 85% (non-condensing) Storage: Upto 95% |
| Ambient Pressure Range | Upto 20 kPa |
Classification & Compliances
| Device Classification | Continuous Operations, Class I, Type B, Drip Proof IPX1; Non-Sterile, Active Medical Device – AC |
| Compliances & Certifications | IEC 60601-1 (Safety); IEC 60601-1-2 (EMC) |
| Intended Patient Population | Neonate, Pediatric and Adult (All age group of patients; requiring non-invasive ventilation) |
Accessories
- Temperature Probe (Standard)
- Heater-Wire Adaptor for Single Limb Disposable Circuit (Optional)
- Heater-Wire Adaptor for Dual Limb Disposable Circuit (Optional)
- Heater-Wire Adaptor for Single Limb Reusable Circuit (Optional)
- Heater-Wire Adaptor for Single Limb Reusable Circuit (Optional)
- Flat Mounting Clamp (Optional)
Note: Product specifications, part / model numbers, and accesories options are subject to change, without prior notice.